Publication:
Ultra-low power digital front-end for single lead ECG acquisition integrated with a time-to-digital converter

dc.contributor.affiliationDA-IICT, Gandhinagar
dc.contributor.authorMishra, Biswajit
dc.contributor.authorThakkar, Sanket
dc.contributor.authorJain, Nupur
dc.contributor.authorMishra, Biswajit
dc.contributor.authorMishra, Biswajit
dc.contributor.authorMishra, Biswajit
dc.contributor.authorMishra, Biswajit
dc.contributor.authorMishra, Biswajit
dc.contributor.researcherThakkar, Sanket (201611046)
dc.contributor.researcherJain, Nupur (201221008)
dc.date.accessioned2025-08-01T13:09:37Z
dc.date.issued16-07-2019
dc.description.abstractA low power single lead electrocardiogram front-end acquisition system in 0.18 ?m CMOS operating at 0.5 V is presented here. The analogue blocks in low noise amplifier (LNA), filters and passive elements that perform amplification and DC offset cancellation are replaced by a moving average voltage to time converter (MA-VTC) to get amplification and anti-aliasing in the time domain. A digital feedback algorithm is used to cancel out the DC offset. The front-end structure is designed in the sub-threshold region of MOS to reduce the power consumption in the circuit. The proposed architecture consumes 50 nW of power with a gain of 670 ?s/V. The output of the front-end is fed to an all digital time-to-digital converter (TDC) that operates in the near threshold region with a resolution of 586.4 ps and 32.5 ?W power consumption.
dc.format.extent453-460
dc.identifier.citationMishra, Biswajit, Sanket Thakkar and Nupur Jain"Ultra-low power digital front-end for single lead ECG acquisition integrated with a time-to-digital converter," IET Computers and Digital Techniques, Institution of Engineering and Technology, vol. 13, issue. 6, ISSN: 17518601, pp. 453-460, 2019, doi: 10.1049/iet-cdt.2019.0027.
dc.identifier.doi10.1049/iet-cdt.2019.0027
dc.identifier.issn1751-861X
dc.identifier.scopus2-s2.0-85074286449
dc.identifier.urihttps://ir.daiict.ac.in/handle/dau.ir/2074
dc.identifier.wosWOS:000493114300005
dc.language.isoen
dc.publisherIET
dc.relation.ispartofseriesVol. 13; No. 6
dc.source IET Computers and Digital Techniques, Institution of Engineering and Technology
dc.source.urihttps://ietresearch.onlinelibrary.wiley.com/doi/10.1049/iet-cdt.2019.0027
dc.titleUltra-low power digital front-end for single lead ECG acquisition integrated with a time-to-digital converter
dspace.entity.typePublication
relation.isAuthorOfPublicationa0d02c67-380f-4408-bf67-20c265f67093
relation.isAuthorOfPublicationa0d02c67-380f-4408-bf67-20c265f67093
relation.isAuthorOfPublication.latestForDiscoverya0d02c67-380f-4408-bf67-20c265f67093

Files

Collections